液體(ti)石(shi)油產品經類測定儀操作(zuo)規程

1.將清洗好的吸(xi)附(fu)柱(zhu)(zhu)末端球形(xing)磨口與(yu)球形(xing)窩之間夾一小(xiao)片(pian)脫脂棉,用小(xiao)鋼夾夾緊。開動(dong)震(zhen)(zhen)動(dong)器,邊(bian)震(zhen)(zhen)動(dong)邊(bian)填(tian)(tian)(tian)充(chong)(chong)硅膠(jiao)。填(tian)(tian)(tian)充(chong)(chong)至分離段一半時,停止(zhi)震(zhen)(zhen)動(dong),填(tian)(tian)(tian)充(chong)(chong)3-5mm熒光指示劑,再填(tian)(tian)(tian)充(chong)(chong)部(bu)分硅膠(jiao),再開啟震(zhen)(zhen)動(dong),使硅膠(jiao)填(tian)(tian)(tian)充(chong)(chong)到(dao)加料(liao)段75mm,再震(zhen)(zhen)動(dong)4min后(hou)停止(zhi)。振動(dong)吸(xi)附(fu)柱(zhu)(zhu)時,用濕布擦拭(shi)整(zheng)個(ge)吸(xi)附(fu)柱(zhu)(zhu),消除(chu)靜電,有助(zhu)于(yu)更好地填(tian)(tian)(tian)充(chong)(chong)硅膠(jiao)。

2.樣品(pin)與取樣器冷卻到4C以下。用取樣器抽取0.75+0.03ml樣品(pin),將(jiang)針頭(tou)插入(ru)硅膠3Omm后(hou)將(jiang)樣品(pin)注入(ru),使樣品(pin)充分吸

附,然后(hou)向(xiang)吸(xi)附柱(zhu)內注入(ru)異(yi)丙醇,至大(da)球窩的下端。

3.試樣(yang)(yang)全部(bu)被(bei)吸附(fu)后,向加料段中注(zhu)入異丙醇至(zhi)球形接頭處(chu),將吸附(fu)柱(zhu)頂(ding)端與供(gong)氣系統相連(lian)接,用夾子夾緊球形磨口接頭。供(gong)氣,在(zai)14kPa±2kPa壓(ya)力(li)下(xia)保持2.5min+0.5min,使液體沿著吸附(fu)柱(zhu)向下(xia)行進。然后加壓(ya)至(zhi)34KPat2kPa,再保持2.5min+0.5min。最后將氣壓(ya)調到適當的壓(ya)力(li),使液體向下(xia)行進的時(shi)間約為1h。通常汽(qi)油類試樣(yang)(yang)大約需要28KPa-69KPa。

4.當紅(hong)色(se)的(de)(de)(de)醇-芳(fang)經界面(mian)(mian)進入分(fen)析段約350mm后,按(an)以下(xia)(xia)順序(從下(xia)(xia)至上(shang))迅速標記(ji)出(chu)(chu)在(zai)紫外燈光(guang)下(xia)(xia)觀(guan)察到的(de)(de)(de)各經類的(de)(de)(de)界面(mian)(mian),測得一(yi)組數據(警告∶直接暴(bao)露在(zai)紫外光(guang)下(xia)(xia)是有(you)害的(de)(de)(de),在(zai)操作(zuo)中宜(yi)盡量(liang)避(bi)免紫外線照射,尤其注意保護眼睛)。對(dui)無熒光(guang)的(de)(de)(de)飽和(he)經區(qu)域(yu),需標記(ji)出(chu)(chu)試(shi)樣(yang)的(de)(de)(de)前沿和(he)黃色(se)熒光(guang)shou次(ci)(ci)達到zui強(qiang)的(de)(de)(de)位(wei)置(zhi);對(dui)于第(di)二部(bu)(bu)(bu)分(fen)即(ji)烯烙(luo)區(qu)域(yu)的(de)(de)(de)上(shang)端,標記(ji)出(chu)(chu)shou次(ci)(ci)出(chu)(chu)現強(qiang)藍色(se)熒光(guang)的(de)(de)(de)位(wei)置(zhi);對(dui)于第(di)三部(bu)(bu)(bu)分(fen)即(ji)芳(fang)經區(qu)域(yu)的(de)(de)(de)上(shang)端,標記(ji)出(chu)(chu)第(di)一(yi)個(ge)(ge)紅(hong)色(se)或(huo)棕(zong)色(se)壞的(de)(de)(de)上(shang)端。對(dui)于無色(se)的(de)(de)(de)餾分(fen),通過一(yi)個(ge)(ge)紅(hong)色(se)環可以清楚地確定醇-芳(fang)經界面(mian)(mian),但裂(lie)化燃料中的(de)(de)(de)雜質常會(hui)使這(zhe)個(ge)(ge)紅(hong)色(se)環變得模糊,出(chu)(chu)現長(chang)度不定的(de)(de)(de)棕(zong)色(se)區(qu)域(yu),它(ta)仍作(zuo)為芳(fang)豎(shu)區(qu)域(yu)的(de)(de)(de)一(yi)部(bu)(bu)(bu)分(fen)來計算。只有(you)在(zai)吸附(fu)柱(zhu)中不出(chu)(chu)現藍色(se)熒光(guang)的(de)(de)(de)情(qing)況下(xia)(xia),此(ci)棕(zong)色(se)或(huo)紅(hong)色(se)環才被(bei)認(ren)作(zuo)是環下(xia)(xia)面(mian)(mian)另(ling)(ling)一(yi)個(ge)(ge)可辨區(qu)域(yu)的(de)(de)(de)一(yi)部(bu)(bu)(bu)分(fen)。對(dui)于某(mou)些含有(you)含氧化合物的(de)(de)(de)調合燃料樣(yang)品,在(zai)紅(hong)色(se)或(huo)棕(zong)色(se)醇-芳(fang)界面(mian)(mian)的(de)(de)(de)上(shang)面(mian)(mian)可能出(chu)(chu)現另(ling)(ling)一(yi)個(ge)(ge)幾厘米長(chang)的(de)(de)(de)

紅色(se)(se)帶(dai),此紅色(se)(se)帶(dai)應忽視,不計入(ru)經(jing)經(jing)色(se)(se)帶(dai)中(zhong)。標(biao)記(ji)各經(jing)類區(qu)域界面時應避(bi)免手與吸附柱表(biao)面接觸(chu)。如果(guo)用(yong)指示夾標(biao)記(ji)各經(jing)類區(qu)域界面,則直(zhi)接記(ji)錄測量(liang)結果(guo)。